Model UT90A: OPERATING MANUAL
20
1O
Open circuit voltage is
about 3V; Set the
resistance of a open
circuit> 100O, buzzer will
not sound; Set the
resistance of a well
connected circuit=10O,
buzzer sounds
continuously.
Remark:
?Overload Protection: 230V rms.
Fuse Replacement(see figure 9)
Warning
To avoid electrical shock or injury to the user
because of wrong readings of the Meter,
please inspect the damage of fuse inside
when there is not any response on the
display during current measurement. Replace
the damaged fuse with identical model
number or electrical specifications.
To replace the fuse as follows:
1. Turn the Meter to OFF position, and remove
the test leads from terminals.
2. Take off the holster as figure 9 shows.
3. Remove the 3 screws from the case bottom,
separate the case bottom, then replace the
damaged fuse.
MAINTENANCE
This section provides basic maintenance
information including battery replacement
instruction.